A functional description
1. How it works: CHWJ - 2024 total arsenic online water quality analyzer measurement method using ISO6595, GB - 7485-87 "water quality - determination of total arsenic - silver diethyl dithiocarbamate spectrophotometry". Zinc and acid, generate new ecological hydrogen in the presence of potassium iodide and stannous chloride pentavalent arsenic reduction for trivalent trivalent arsenic is nascent hydrogen reduction into hydrogen arsenide (arsine) using silver diethyl dithiocarbamate triethanolamine chloroform liquid absorption arsine generate colloidal silver in 530 nm wavelength red absorption solution's absorbance measurement. Instrument will directly by color conversion total arsenic value displayed in the samples.
- display the time can be set, proofreading, including year, month, day, and when,;
- with a power device, leakage of the instrument, sample can't import the reactor system alarm and display abnormal conditions such as fault. Stop running until the system is restarted at the same time;
- protection of cutting off power and water and electricity, the coming water automatic recovery function;
- do regular cleaning, regular sample function;
- touch screen to view more intuitive, more convenient operation;
Test range: 0-3 mg/L (can expand according to the requirement)
Detection limit: 0.001 mg/L
Resolution: 0.001 mg/L
The accuracy of the quasi: no more than 10% of error
Heavy after sex: the relative standard deviation less than 5%
Zero drift, plus or minus 5% or less
Span drift: plus or minus 10% or less
Few reagent consumption, accurate measurement, reagent dosage, cost savings for the user
User maintenance: maintenance interval > 1 month, about 1 hour each month; Maintenance is simple and convenient
Environmental impact: extreme water quality (such as high suspended solids, impurities, floaters) without damage
Communication interface: RS232 digital interface, 4 ~ 20 ma analog output
Installation dimensions: 1400 * 560 * 450 mm deep
Weight: 65 kg
Electricity source, AC (220 + / - 10%) V and Hz (50 + / - 10%), 5 a
Ambient temperature: 5 ~ 40 degrees Celsius
Condensation environment humidity: 90% or less (not)
The atmospheric pressure: 86000 mpa to 106000 mpa
Sampling PaiKou: distance L 15 m or less, divide H 6 m or less